M500T Multi-parameter Analyzer

500.00৳ 

⛟ Stock products delivery time 1-3 working days. 🔥
✈ Out-of-stock products delivery time 10-20 working days. 🚀

☎️ Contact Us for any inquiry

Description

M500T

GENERAL FEATURES

● Color High contrast LCD touchscreen, 7.0 inches.

● Intelligent operation system provides features including user management, calibration management, method management, data management and etc.

● Multi-reading feature allows auto-read, timed-read and continuous-read.

● Automatic/Manual temperature compensation ensures accurate results.

● Data Storage 1000 sets for each parameter (GLP-compliant).

● Data analysis feature helps user review, compare and recalculate results.

● Support for USB or RS-232 communication.

● Reset feature automatically resumes all settings back to factory default options.

● IP54 waterproof.

Large color touch screen

Large color touch screen

Easy operation and comfortable experience

Autosampler available

Autosampler available

Make measurement more  efficient

User management system

User management system

Multiple external devices available

Multiple external devices available

PC, autosampler, printer, scanning  gun, USB flash disk, etc

pH

● 1-6 points calibration with Standard Recognition.

● Selectable pH buffer groups, including DIN, NIST, USA, MERK, JIS, GB.

● Support for pH standard group customization.

● Automatic electrode diagnosis with pH slope and offset display.

 

Ion

● 1-6 points calibration

● Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, ppm, etc.

● Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.

● Over 10 methods are built-in, including F- , Cl- , Br- , I- , NO3 – , BF4 – , NH4 + , K+ , Na+ , Ca2+, Cu2+, Pb2+, Ag+ and etc., user-defined method is supported.

 

Conductivity

● 1-5 points calibration with Standard Recognition.

● Settable parameters, including cell constant, reference temperature (5/10/15/18/20/25°C), temperature compensation coefficient and TDS factor.

● Temperature compensation type (none, linear, pure water).

 

DO

● Support for air-saturated water or zero oxygen calibration.

● Auto barometric pressure compensation.

● Manual Salinity Factor Correction.

● Selectable pressure unit, including kPa, mbar, Torr, Atm.

Specification

Model M500T
Parameters pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal./DO Saturation)
              pH Range -2 .000 to 20.000 pH
Resolution 0.1, 0.01, 0.001pH
Accuracy ± 0 .002 pH
Calibration Points Up to 6
Standard Customization Yes
Calibration Reminder Yes
Standard Recognition NIST, GB, DIN, USA, MERK, JIS buffers
Slope Limit Yes
         mV/ORP Range -2000 .00 to 2000.00 mV
Resolution 0.1, 0.01 mV
Accuracy ± 0 . 1 mV or ± 0 .03% of reading whichever is greater
EH ORP Mode Yes
ORP Calibration Points 1 custom point (relative mV)
                pX Range -2 .000 to 20.000 pX
Resolution 0.1, 0.01, 0.001 pX
Accuracy ± 0 .002 pX
Calibration Points Up to 6
           ISE Range 1E-9 to 9 .999E9
Units mol/L, mmol/L, g/L, mg/L, μ g/L, ppm
Resolution Up to 4 significant digits
Accuracy ± 0 .3%
Calibration Points Up to 6
       Conductivity Range 0.000 μ S/cm to 2000 mS/cm
Resolution 0.001 μ S/cm minimum,

various with range selection

Accuracy ± 0 .5% FS
Reference Temperature 5, 10, 15, 18, 20, 25 ºC
Calibration Points Up to 5
Calibration Reminder Yes
Standard Recognition 10 μ S/cm, 84 μ S/cm, 500 μ S/cm, 1413 μ S/cm, 12.88 mS/cm
Resistivity Range 5.00 Ω ·cm~100.00 M Ω ·cm
Resolution 0.01 Ω ·cm minimum
Accuracy ± 0 .5% FS
TDS Range 0 .000~1000 ppt
Resolution 0.001 ppm minimum, changed with range
Accuracy ± 0 .5%FS
 

Salinity

Type NaCl%
Range (0.00~8.00)%
Resolution 0.01%
Accuracy ± 0 . 1%
 Dissolved Oxygen

Concentration -Polarographic

Range 0.00 to 99.99 mg/L
Resolution 0.01 mg/L
Accuracy ± 0.3mg/L (0.00 to 20.00 mg/L);

± 10% (20.00 to 99.99 mg/L)

Calibration Reminder Yes
Calibration Points Air-saturated water or zero point
Barometric Compensation Yes
Manual Salinity Factor Correction Yes
% Saturation -Polaro- graphic Range (0 .0 to 600)%
Resolution 0.10%
Accuracy ± 2 .0%
         Temperature Range – 10 to 135 ºC, 14 to 275 ºF
Unit ºC, ºF
Resolution 0.1
Accuracy ± 0 . 1
Measurement Reading Mode Auto Read (Fast, Medium, Slow), Timed, Continuous
Reading Prompts Reading, Stable, Locked
Temp. Compensation ATC, MTC
Data management Data Storage 1000 Groups
GLP Features Yes
Log Management Yes
Inputs pH Electrode BNC(Q9)
DO with Temp . Probe 4-pin aviation connector
Conductivity with Temp . Probe 5-pin aviation connector
Outputs USB USB 2.0 flash memory device, PC
RS 232 printer, scanner, auto-sampler
Shipping & Returns Policy

Shipping Information

Dhaka Only: Same-day or next-day delivery within Dhaka city limits and cost only ৳80 TAKA.
Out Of Dhaka: 1–3 working days for most locations within Bangladesh and cost only ৳120 TAKA.

Know more details. 

Return Policy

You may return items purchased from Meter BD Shop within 7 days of receiving the product.

To be eligible for a return, the product must meet the following conditions:
1. Unused, 2. In the original packaging & 3. Proof of purchase.

Read More

Reviews

There are no reviews yet.

Be the first to review “M500T Multi-parameter Analyzer”

Your email address will not be published. Required fields are marked *