NADE SYD-261A Automatic Pensky-Martens Closed Cup Flash Point Tester & Fire Point Tester for Petroleum Products ASTM D93
500.00৳
⛟ Stock products delivery time 1-3 working days. 🔥
✈ Out-of-stock products delivery time 10-20 working days. 🚀
☎️ Contact Us for any inquiry
NADE SYD-261A Pensky-Martens Closed Cup Flash Point Tester & Fire Point Tester for Petroleum Products ASTM D93
This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point – Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.
1. Simulation tracking displays function curves of heating up and test time.
2. Equipped with the English prompt function of mis-operation.
3. Equipped with prompt function of test date,test time and other parameters.
4. Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.
5. Differential coefficient detection.Automatically correct the system deviation.
6. Automatically open the lid,detect the flash point and print test data.
7. The test arm rises up and lowers down automatically.
8. Igniting with electric igniting gun,forced air cooling.
1. Power supply: AC(220±10%)V, 50Hz.
2. Flash point determination:Range: Ambient to 300℃.
Repeatability: ≤ 2℃
Reproducibility: ≤ 4℃
Accuracy: 0.1℃
3. Heating rate: Procedure A: (5~6)℃/min
Procedure B:(1~1.6)℃/min
Automatic control and manually adjustable.
4. Stirring rate:Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5. Igniting mode: (1) Igniting source: gas(or other civilian fuels,the same below)
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
6. Ambient temperature: ≤ 40℃
7. Relative humidity: ≤ 80%
8. Maximum power consumption: 500W
Shipping Information
Out Of Dhaka: 1–3 working days for most locations within Bangladesh and cost only ৳120 TAKA.
Know more details.
Return Policy
You may return items purchased from Meter BD Shop within 7 days of receiving the product.
To be eligible for a return, the product must meet the following conditions:
1. Unused, 2. In the original packaging & 3. Proof of purchase.
Reviews
There are no reviews yet.