Standard Analyzer SCP Panel, Medium Pressure, ˂600 psig (41.4 bar), ˂600 °F (315 °C)

⛟ Stock products delivery time 1-3 working days. 🔥
✈ Out-of-stock products delivery time 10-20 working days. 🚀

☎️ Contact Us for any inquiry

Description

Simple to install. Safe to operate.

Hach’s Sample Conditioning Panel (SCP) for a single analyzer allows any analyzer to be easily connected to a hot and pressurized sample. It helps to protect and respect all safety and sample requirements.

Applications:

  • Boiler Water
  • Steam Cycle
  • Water Cycle

Benefits

Complete Solution

Everything from one single source: pre-configured panel, service, digital learning modules, and quality control. The single standardized platform minimizes time required to teach and learn product operations, getting new systems in use faster.

Easy and Safe Operation

The SCP (Sample Conditioning Panel) is easy to install and requires very little maintenance. Safe operation through shut-off valves for hot samples, safety valves for cooling water, protection for hot parts, and certified pressure tests.

Robust and Industrial Design

The efficient cooler designs for each panel ensures the need for less cooling water for the wide sample pressure and temperature ranges. Key parts are in stainless steel. The design allows for easy connection to any single independent analyzer or for integration in any complete skid of analyzers.

Specifications

Technical Attributes

Application Make up water, feed water, boiler water, condensate and steam in medium pressure boilers for Power & Industry
Cooling Water Filtered, softened/decarbonized water quality required
Permissible Chloride Range <250 mg/L for sample temperature 77 – 356 °F (25 – 180 °C)
<100 mg/L for sample temperature 356 – 554 °F (180 – 290 °C)
<25 mg/L for sample temperature 554 – 1022 °F (290 – 550 °C)
pH Range 7 – 12 pH
Pressure Range Cooling water: <150 psig (10.3 bar)
Sample Flow Rate 200 mL/min for conductivity
150 mL/min for SiO₂ and PO₄
100 mL/min for pH, O₂, Na
Plus 350 mL/min for grab sample

Maximum total sample flow for the panel is 1600 mL/min for hot water or 1000 mL/min for steam.
Up to 3 analyzers can be connected to one sampling panel.

Sample Pressure Liquid: 50 – 500 psig (3.4 – 34.4 bar)
Steam: 300 – 500 psig (20.7 – 34.4 bar)
Sample Temperature <600 °F (315 °C)
Temperature Cooling water: <110 °F (43 °C)
Turbidity <50 NTU
What’s included? Sample Conditioning Panel for online analysis, consisting of the following components:
Sentry Back Pressure Relief Valve with a Set-point of 20 psig (1.4 bar), Grab Sample, 304 SS Sample Sink; all components are factory mounted on a 12 in. (W) x 36 in. (H), T304SS backplate. All sample tubing throughout panel is SA249, T316 SS.
Cooling Water Valve Materials are Stainless Steel. Sentry Thermal Shut off Valve with a Set-point of 120 °F (49 °C).
Low Pressure (Cooled Sample) Blow-down Valve, Total Flow Indicator; 0-1600 cc/min, Temperature Gauge; 0-250 °F (-20-120 °C) Pressure Gauge; 0-60 psig (0-4 bar). Globe Valve on the Cooling Water Outlet. 3-Way Cooling Water Inlet Valve with drain.
1/4 in. Sample Inlet Isolation Valve. Primary Sample Cooler, Sentry Model TLR4225. Sample Flow Control Valve.
Shipping & Returns Policy

Shipping Information

Dhaka Only: Same-day or next-day delivery within Dhaka city limits and cost only ৳80 TAKA.
Out Of Dhaka: 1–3 working days for most locations within Bangladesh and cost only ৳120 TAKA.

Know more details. 

Return Policy

You may return items purchased from Meter BD Shop within 7 days of receiving the product.

To be eligible for a return, the product must meet the following conditions:
1. Unused, 2. In the original packaging & 3. Proof of purchase.

Read More

Reviews

There are no reviews yet.

Be the first to review “Standard Analyzer SCP Panel, Medium Pressure, ˂600 psig (41.4 bar), ˂600 °F (315 °C)”

Your email address will not be published. Required fields are marked *